<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-3
IM 01E30D01-01EN
NOTE
Before updating any setting, remember to always
check the parameter content you want to change
as described in Chapter 6.
NOTE
IncaseofHARTConguration,theparameters
are displayed in English only.
Even if the language with the exception of
English is selected at “Language” setting display,
the parameters are displayed in English upon
HARTCongurationTool.
9.4 Basic Setup
ConrmationandChageofTagand
Device Information
The tag number and device information can be
checked as follows:
• Thelocationforthetagnumberanddevice
information
Tag [Root Menu]→DetailedSetup
→Basicsetup→Tag
Long tag [Root Menu]→DetailedSetup
→Basicsetup→Long tag
Descriptor [Root Menu]→DetailedSetup
→Deviceinfo→Fielddeviceinformation
→Descriptor
Message [Root Menu]→DetailedSetup
→Deviceinformation
→Fielddeviceinformation→Message
Date [Root Menu]→DetailedSetup
→Deviceinfo→Fielddeviceinformation
→Date
When changing the tag number or device
information, enter the information directly within the
following limitations.
Item Limitations
Tag Up to 8 characters or numbers*
1
Long tag
Up to 32 characters or numbers*
2
Descriptor Up to 16 characters or numbers*
1
Message Up to 32 characters or numbers*
1
Date mm/dd/yyyy (DD)
yyyy/mm/dd (DTM)
- mm: month (2 digits)
- dd: days (2 digits)
- yyyy: years (4 digits)
9.5 ParametersConguration
TheparameterstructureoftheHARTconguration
tool is hierarchical.
Read Section 9.9, Menu Tree Example for the
corresponding parameters. The menu tree shows
a cross-reference of the parameters for HART and
BRAIN.
Read Chapter 6 for the functions of parameters.
NotethatsomedisplayparametersofAXRare
differentfromthoseofHARTcongurationtools.
9.6 Data Renewing
TherearetwomethodstoloadtheAXRdata
from/toHARTCongurationTool---periodicdata
renewing and discretionary data renewing. Read
Section 9.9 menu tree for each corresponding
parameter.
(1) Periodic Data Renewing
The data is updated 0.5 to 2 second cycles. The
parameter of this type is marked as “D” in “Data
Renewing” of Section 9.9 menu tree.
(2) Discretionary Data Renewing
Thedatacanbeloadedfrom/totheAXRwhen
thedatasaveisnishedtotheConguration
Tool. The parameter of this type is marked as
“S” in “Data Renewing” of Section 9.9 menu
tree.
NOTE
TheAXRisnotabletosetparametersviaits
display unit during a communication of the HART
CongurationTool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals