EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ADMAG AXR

YOKOGAWA ADMAG AXR User Manual

YOKOGAWA ADMAG AXR
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
<14. EXPLOSION PROTECTED TYPE INSTRUMENT>
14-5
IM 01E30D01-01EN
SpecicationofProtection:
Electrode Circuit Um: 250 V
Power Supply/Current Output: 42Vdc max./
4 to 20mA
Digital Output:
On: 2Vdc, 120mA max.
Off: 30Vdc max., 4mA
Excitation Circuit: 29V max.
Enclosure: IP66/IP67
Process Temperature:
Temperature
Code
Maximum Process
Temperature
Minimum Process
Temperature
T6 +70°C (+158°F) -40°C (-40°F)
T5 +85°C (+185°F) -40°C (-40°F)
T4 +130°C (+266°F) -40°C (-40°F)
Ambient Temperature: -40 to +55°C (-40°F to
+131°F)
Type of Dust Atmosphere Protection
Type of Protection:
DIP A21 T
A
90°C, 110°C, 130°C
SpecicationofProtection:
Electrode Circuit Um: 250 V
Power Supply/Current Output: 42Vdc max./
4 to 20mA
Digital Output:
On: 2Vdc, 120mA max.
Off: 30Vdc max., 4mA
Excitation Circuit: 29V max.
Enclosure: IP66/IP67
Process Temperature:
Maximum Surface
Temperature T
A
Maximum Process
Temperature
Minimum Process
Temperature
90°C +70°C (+158°F) -40°C (-40°F)
110°C +85°C (+185°F) -40°C (-40°F)
130°C +130°C (+266°F) -40°C (-40°F)
Ambient Temperature: -40 to +55°C (-40°F to
+131°F)
ProcessSealingCertication:
DualSealcertiedbyCSAtotherequirements
of ANSI/ISA 12.27.01-2003.
No additional sealing required.
Primary seal failure annunciation;
Deteriorationoftheowrateoutputat
nonzeroowpoint.
Unstableowrateoutputatzeroowpoint.
(2) Installation
For Division
WARNING
• AllwiringshallcomplywithCanadian
Electrical Code Part I and Local Electrical
Codes.
• Inhazardouslocation,wiringshallbein
conduit as shown in Figure 14.3.
WARNING : A SEAL SHALL BE INSTALLED
WITHIN 50cm OF THE
ENCLOSURE.
UN SCELLEMENT DOIT ÊTRE
INSTALLÉ Á MOINS DE 50cm DU
BOÎTIER.
• WheninstalledinDivision2,“SEALSNOT
REQUIRED”
Conduit
Magnetic Flowmeter
Sealing Fitting
HAZARDOUS LOCATIONS
50 cm Max.
F1404.ai
Figure 14.3 Conduit Wiring
For Zone
WARNING
• AllwiringshallcomplywithCanadian
Electrical Code Part I and Local Electrical
Codes.
• Inhazardouslocations,thecableentry
devicesshallbeofacertiedameproof
type, suitable for the conditions of use and
correctly installed.
• Unusedaperturesshallbeclosedwith
suitableameproofcertiedblanking
elements.(Theplugattachediscertiedas
theameproofandIP66orIP67asapartof
this apparatus.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals