101510 Rev. AG Page 3–
55
3.28 Sealing Environmental Enclosure
When all sample and vent tubing has been installed and
leak-tested, the sample boot must be sealed. The following
procedures describe the steps for sealing the environmental
enclosure.
3.28.1 Customer Supplied Materials
Aerosol insulating foam
3.28.2 Instructions
1) When all sample and vent connections are complete, apply aerosol
insulating foam from inside the enclosure pointing toward the outside of
boot, ensuring that the overspray falls outside the enclosure.
3.29 Optional Equipment Enclosure Installation
If the optional enclosure is used, it may be configured to include other options,
including, but not limited to, a battery pack to provide reserve power to the NGC,
communication equipment, solar power charger and additional I/O.
Three enclosures are commonly used for the NGC8206 installations: the 6200,
6700 and 6800 enclosure. The unit may be mounted on a 2” pipe or mounted on a
flat surface such as a wall.
The battery and solar panel are packed and shipped separately from the
enclosure.
Before beginning, review the procedures and the materials required for the
installation; inspect all power cables where they terminate and the connector for
breakage.
The optional equipment enclosure may be approved for
classified hazardous locations or potentially explosive
atmospheres. Verify the rating listed on the unit tag and install
per the referenced control drawing. Be sure to follow the
requirements of national and local codes when installing the
optional equipment enclosure.
3.29.1 6200 Optional Equipment Enclosure
The 6200 can accommodate the following equipment:
Power Supply Kit for 6200
120 VAC/12 VDC power supply
240 VAC/12 VDC power supply
The 6200 installation will be for AC and 24 VDC sites. There is no battery-backed
option in this installation.
3.29.2 6700 Optional Equipment Enclosure
The 6700 optional equipment enclosure can accommodate the following:
Power Supply Kit
120/240 VAC / 12 VDC power supply
24 VDC/12 VDC DC to DC converter
ABB Totalflow XFC/XRC electronic board
XFC/XRC onboard battery charger
Battery Options

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB NGC8206 and is the answer not in the manual?

ABB NGC8206 Specifications

General IconGeneral
BrandABB
ModelNGC8206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals