Accu-Chek D-TRONplus insulin pump Reference Manual
101
Accu-Chek D-TRONplus insulin pump Reference Manual
100
5.3 Change the cartridge, adapter and infusion set
Warning: Never change the cartridge or prime the infusion set with an
infusion set inserted into your body. You risk uncontrolled insulin
delivery. With disconnectable infusion sets, make sure you have
disconnected prior to changing or priming.
Always follow the instructions for use of the infusion set you are using.
Caution: Air bubbles in the cartridge, the adapter and in the infusion
set cause the infusion of air instead of insulin. Your body does not receive
the required amount of insulin. An “Error E4: Occlusion” may be delayed.
Remove these bubbles while filling the cartridge and/or priming the infu-
sion set. Make sure the infusion set is not connected to your body.
Cold insulin may release air when warmed. Use only insulin at room tem-
perature when filling the cartridge and priming the infusion set.
Inspect the cartridge, adapter and the infusion set for air bubbles at least
every three hours during the day and before you go to sleep. Remove any
air bubbles and, if necessary, change system components.
Caution: Always handle the items with clean hands.To make sure that the
cartridge compartment is always clean, change the cartridge and adapter
in clean environment and check always that the cartridge compartment is
free of any dirty particles. Avoid your insulin pump, the infusion set and es-
pecially the connecting parts of your insulin pump coming into contact
with health and beauty items (e.g. antiseptics, antibiotic creams, soaps,
perfumes, deodorants, body lotions or any other cosmetics). They may
cause damage to these parts.
To cancel an extended bolus
During programming:
If you do not press any key for 20 seconds*, your insulin pump
automatically returns to RUN (timeout).
No bolus is delivered!
Set the bolus amount to 0.0 units by pressing/scrolling
s
.
Press
f
and your insulin pump returns to RUN.
No bolus is delivered!
During bolus delivery:
If the extended bolus delivery has begun, it can be cancelled by
putting your insulin pump into STOP.
This cancels the delivery of the bolus and an Alert A8: Bolus
Cancelled” occurs.
Silence and confirm the alert by pressing
f
twice and put your
insulin pump back into RUN (press
d
and
f
).
Please refer to section 7.1.8 Alert A8: Bolus Cancelled” for fur-
ther information.
The bolus amount delivered before the cancellation can be reviewed in the bolus his-
tory. Please refer to section 6.1.1 Review the bolus history” for further information.
Caution: After an extended bolus has been cancelled, your insulin pump
is in STOP. Return your insulin pump to RUN to continue insulin delivery.
* These parameters may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek D-TRONplus and is the answer not in the manual?

Accu-Chek D-TRONplus Specifications

General IconGeneral
BrandAccu-Chek
ModelD-TRONplus
CategoryMedical Equipment
LanguageEnglish

Related product manuals