Accu-Chek D-TRONplus insulin pump Reference Manual
139
Accu-Chek D-TRONplus insulin pump Reference Manual
138
Set up your insulin pump with a new cartridge, adapter and
infusion set tubing and prime.
Please refer to section 4 Prepare the cartridge and attach the
adapter and infusion set” for further information.
When finished, reconnect the infusion set tubing and put your
insulin pump into RUN (press
d
and
f
).
5.
If the adapter is connected incorrectly (with inserted car-
tridge):
Remove the adapter and infusion set tubing from
your insulin pump.
Set up your insulin pump with a new adapter and infusion set
tubing and prime.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information.
When finished, reconnect the infusion set tubing and put your
insulin pump into RUN (press
d
and
f
).
6.
If the piston rod is accidentally returned or not in contact
with the cartridge plunger (with correctly connected
adapter):
Prime until the piston rod reaches the cartridge
plunger, reconnect your infusion set tubing and put your in-
sulin pump into RUN (press
d
and
f
).
7.
If the cartridge is leaking (with piston rod in position):
Remove the cartridge, adapter and infusion set tubing from
your insulin pump.
If the cartridge is broken or leaking, clean and dry the car-
tridge compartment of your insulin pump.
Please refer to section 10.2.2.“Your insulin pump” for further
information.
Set up your insulin pump with a new cartridge, adapter and
infusion set tubing and prime.
Please refer to section 4 Prepare the cartridge and attach the
adapter and infusion set” for further information.
When finished, reconnect the infusion set tubing and put your
insulin pump into RUN (press
d
and
f
).
7.1.4 Alert A4: Cartridge/Adapter Alert
The unique force sensor of your Accu-Chek D-TRONplus insulin pump continuously
measures the pressure on the cartridge plunger. This feature determines whether a
cartridge is present in the cartridge compartment and whether the adapter is prop-
erly connected.The adapter itself features a valve which prevents uncontrolled deliv-
ery of insulin. If this valve is damaged, an Alert A4: Cartridge/Adapter Alert”
occurs.
If your insulin pump undergoes a thermal shock, that is to say a rapid major decrease
of temperature, an “Alert A4: Cartridge/Adapter Alert”occurs.
If the insulin delivery is impaired by a leak in the cartridge or in the adapter, anAlert
A4: Cartridge/Adapter Alert”occurs.
This alert also occurs when your insulin pump is started and no cartridge is in the car-
tridge compartment, when the adapter is connected incorrectly, the piston rod is not
in contact with the cartridge plunger or the piston rod is accidentally returned.
Caution: If an Alert A4: Cartridge/Adapter Alert” occurs, immediately
check your blood glucose level, because insulin delivery may have been
interrupted. If your blood glucose level is high, take the appropriate ac-
tions according to your health care team’s instructions.
With a disconnectable infusion set do the following:
1. Silence and confirm the alert by pressing
f
twice.
2. Check your blood glucose level. If it is high, take the
appropriate actions according to your health care
team’s instructions.
3. Disconnect the infusion set from the site.
4.
If no cartridge is in the cartridge compartment:
Remove the adapter and infusion set tubing from your insulin
pump.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek D-TRONplus and is the answer not in the manual?

Accu-Chek D-TRONplus Specifications

General IconGeneral
BrandAccu-Chek
ModelD-TRONplus
CategoryMedical Equipment
LanguageEnglish

Related product manuals