Accu-Chek D-TRONplus insulin pump Reference Manual
151
Accu-Chek D-TRONplus insulin pump Reference Manual
150
Do the following:
1. Silence and confirm the error by pressing
f
twice.
2. Remove or disconnect your infusion set from the site.
3. Remove the cartridge, adapter and infusion set (tubing)
from your insulin pump.
4. Return the piston rod.
Make sure that the piston rod has returned completely.
If it has not, perform this function again.
5. Set up your insulin pump with a new cartridge, adapter and
infusion set (tubing).
Please refer to section 4 “Prepare the cartridge and attach
the adapter and infusion set” for further information.
6. When finished, (reconnect the infusion set tubing and) put
your insulin pump into RUN (press
d
and
f
).
Repeat these steps if “Error E6: Mechanical Error” occurs again. Should “Error E6:
Mechanical Error”occur a third time in spite of these actions, contact your doctor or
health care team for an alternate therapy plan. Contact your Accu-Chek customer
service center for insulin pumps for further assistance.
7.2.7 Error E7: Electronic Error
The safety system of your insulin pump continuously monitors the functioning of your
insulin pump. If an error is discovered, the insulin delivery is stopped and an “Error
E7: Electronic Error”occurs.
If the display is damaged, it may not show the error code, however, your insulin pump
beeps and vibrates.
7.2.5 Error E5: End of Use Time
An “Alert A5: End of Use Time Alert”occurs 56 days before the End of Use Time.
When the remaining time of use has reached zero, an “Error E5: End of Use Time”
occurs and your insulin pump will go into STOP. It can no longer be put into RUN.
Please refer to section 6.9 “Review the time remaining” for further information.
Do the following:
1. Silence and confirm the error by pressing
f
twice.
Your insulin pump is in STOP. The symbol for the timer t1 or
t2 appears on the display.
2. Remove or disconnect your infusion set from the site.
3. Remove the cartridge, adapter, infusion set and PowerPack
from your insulin pump.
4. Contact your health care team for an alternate therapy plan.
5. Contact your Accu-Chek customer service center for insulin
pumps for further assistance.
7.2.6 Error E6: Mechanical Error
With every insulin delivery (every 3 minutes), with every change of operating status
and with each return of the piston rod, the safety system of your insulin pump re-
views all mechanical parameters. If a mechanical error is discovered during this
process, the insulin delivery is stopped and an “Error E6: Mechanical Error” occurs.