Accu-Chek D-TRONplus insulin pump Reference Manual
111
Accu-Chek D-TRONplus insulin pump Reference Manual
110
3. Press
s
to review the previous bolus.
4. Use
a
or
s
to review further or previous boluses.
If there are less than 10 boluses in the bolus history, there
are dashes in the display.
5. Press
f
twice to return to RUN, or press
d
6 times,
to directly review the history of daily totals.
Info: If you do not press any key for 20 seconds*, your insulin pump auto-
matically returns to RUN (timeout).
* This parameter may vary based on the standard configuration of your insulin pump
and changes made using the Accu-Chek DiaLog Pump Programming tool.
6.1 Review the data memory
The data memory of your Accu-Chek D-TRONplus insulin pump enables you to review
specific events such as:
1. Bolus history last 10 boluses
2. History of daily totals last 7 daily totals
3. History of alerts and errors last 10 alerts and errors
4. Event history last 1400 events
The last 1400 events (representing about 90 days of continuous normal use) are
stored in the memory of your insulin pump. The data memory can be transferred to a
PC using Accu-Chek DiaLog or other software packages. The 1400 events cannot be
displayed on the display of the insulin pump.
6.1.1 Review the bolus history
The bolus history enables you to review the last 10 boluses delivered starting with
the newest entry in chronological order.
To review the bolus history
1. Make sure your insulin pump is in RUN and the KeyLock is
deactivated or the keys are unlocked.
2. Press
d
once to review the last bolus delivered with bolus
amount, time and date of delivery.
In case of an extended bolus, the time when the bolus deli-
very finished is displayed.