Installing the G1330B Thermostat
Controller Requirements
Controller Requirements
Agilent 1100 Series control module and Agilent ChemStation must have the
latest revision firmware/software loaded to allow optimum operation with
the thermostattable autosamplers. Older revisions might not recognize or
have full functionality with the thermostattable autosampler.
Control Module Firmware requirements
The control module requires firmware revision A.01.30 or higher to control
the thermostatted autosampler. Previous firmware revisions of the control
module will not run with the thermostatted autosampler. If the control
module was shipped together with the thermostatted autosampler the
control module firmware does not require updating. The firmware update
must be done with a PCMCIA card that has the newest revision loaded. The
firmware is not part of the shipment of the thermostatted autosampler.
Contact your local Agilent Technologies sales and service office for the
firmware update of the control module. For the update procedure see “Stage
7: Update of Control Module Firmware” on page 37.
Agilent ChemStation Software requirements
To control the thermostatted autosampler from a PC, the Agilent
ChemStation software version A.04.03, or A.05.02beta, or A.05.02 or higher is
required. These software revision might however not support all ALS
modules. The thermostatted autosampler will not run with any previous
version of the Agilent ChemStation software. The software updates are part
of the thermostatted autosampler shipment. For the update procedure see
“Stage 8: Update of the Agilent ChemStation Software” on page 38.