Troubleshooting and Test Functions
Lost Contact to ALS Thermostat
Lost Contact to ALS Thermostat
The autosampler control electronics continuously checks whether the ALS
thermostat is active or not.
Probable Causes: Bad cable connection between autosampler and ALS thermostat
Cable between autosampler and thermostat module not connected
Defective electronic board in autosampler or ALS thermostat
Suggested Actions: Check cable connection between autosampler and ALS thermostat
Connect cable between autosampler and ALS thermostat
Exchange ASM board in the autosampler
Exchange TCA board in the thermostat
contact the Agilent Technologies service organization