Repairing the ALS Thermostat Module
Exchanging the Sensors
Exchanging the Sensors
When required Sensors defective
Preparations for this
procedure
Removing the Top Cover and Foam (see Removing the Top Cover and Foam on page 61).
Parts required Sensor flexboard, PN G1330-66504
Heatsink Compound, PN 6040-0454
1 Unscrew the six holding screws on top of the
hardfoam encapsulation.
2 Carefully lift the upper part of hardfoam
encapsulation away.
Screws