ATI Q46D Dissolved Oxygen System Part 1 - Introduction
6
O&M Manual
Rev-N (4/21)
1.4 Q46D/60 System Specifications
Displayed Parameters Main input, 0.1 ppm to 40.0 ppm
%Saturation, 0 to 999.9%
Sensor temperature, -10.0 to 50.0°C (23 to 122ºF)
Sensor signal, -40 to +2000 mVDC
Loop current, 4.00 to 20.00 mA
Sensor slope/offset
Model number and software version
PID Controller Status
Main Parameter Ranges Manual selection of one of the following display ranges,
0.00 to 40.00 ppm or 0.00 to 40.00 mg/l
0.0 to 999.9% Saturation
Power 90 - 260 VAC, 50/60 Hz., 10 VA Maximum
12-24 VDC, 500 mA max.
Display 0.75” (19.1 mm) high 4-digit main display with sign
12-digit secondary display, 0.3" (7.6 mm) 5x7 dot matrix.
Integral LED back-light for visibility in the dark.
Enclosure NEMA 4X, polycarbonate, stainless steel hardware,
Mounting Options Wall, pipe, or panel mount standard. Wall bracket suitable for
either 1.5” or 2” I.D. U-Bolts for pipe mounting.
Conduit Openings Five ½” NPT openings, Adapter can be removed to provide a 1”
NPT opening in the bottom of the enclosure. Gland seals
provided but not installed.
Relays, Electromechanical: Three SPDT, 6 amp @ 250 VAC, 5 amp @ 24 VDC contacts.
Software selection for setpoint, phase, delay, deadband, hi-lo
alarm, and failsafe. A-B indicators on main LCD, and C indicator
on lower display.
Analog Outputs Two 4-20 mA outputs. Output one programmable for PPM D.O.
or PID. Output 2 programmable for PPM D.O. or Temperature.
Max load 450 Ohms for output 1 and 1000 ohms for output 2.
Outputs ground isolated and isolated from each other. An
additional 3
rd
analog option is available.
Output Isolation 600 V galvanic isolation
Optional Relays: Three SPST, 1 amp @ 24 VDC. Software selection for setpoint,
phase, delay, deadband, hi-lo alarm, and failsafe.
Ambient Temperature Analyzer Service, -20 to 60 °C (-4 to 140 ºF)
Sensor Service, -5 to 55°C (23 to 131 °F)
Storage, -30 to 70 °C (-22 to 158 ºF)
Keypad 4-key membrane type with tactile feedback, polycarbonate with
UV coating

Table of Contents

Other manuals for ATI Technologies Q46 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ATI Technologies Q46 Series and is the answer not in the manual?

ATI Technologies Q46 Series Specifications

General IconGeneral
BrandATI Technologies
ModelQ46 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals