10
PATIENT SAFETY
• While using TCI an appropriate patient monitoring is mandatory.
•
Take care of using the right dilution/concentration of the drug and make sure
the right dilution is selected at the pump.
•
Never administer Propofol or Remifentanil by a second infusion as long as
you use TCI.
•
It is possible to completely switch off the TCI mode to avoid the use of TCI
accidentally.
Safety Instructions for using Pole Clamp
1. Line pump up with the Pole Clamp
guide rails.
2. Slide pump fully into place onto the
guide rails.
3. An audible “Click” should heard.
4. Test the pump is secure.
W
The pump ist now securely attached to
Pole Clamp.
• Do not lean on the pump when
attached to the Pole Clamp.
• Do not position the pump unit
above the patient.
W
• DO NOT use any Pole Clamp that
shows signs of damage.
• DO NOT use Pole Clamp with
missing clamp grids.
Underside view
Clamp grids