6.3 Reminder Alarms
Reminder alarms only occur in two cases:
1. A syringe is inserted, the pump doesn’t administrate, no value is being edited and
the device is not operated for two minutes.
An acoustic tone sounds, the yellow LED is constantly on and a staff call is
activated.
a) The display states “Reminder alarm!
b) The display states “Config. not finished!
Confirm alarm with
K and continue to set therapy/Start Up configuration.
2. A value edition was started but not finished and confirmed. This is also possible
with a missing disposable.
An acoustic tone sounds, the display states Value not accepted“, the yellow LED
is constantly on and a staff call is activated.
Confirm alarm with
K and continue to set therapy.
6.4 Alarm Hints
If inproper entries are made the display states corresponding hints (e.g. Bol.rate out of
range”;Download failed”; The parameter can not be modified”). These hints
disappear after a few seconds and don't need to be confirmed.