ADVANCED OPERATIONS
2.1 Status Request of Pump when Infusion is Running
Press c to switch between run display and Main Menu while the device is infusing.
Navigate through the menu using
t to check parameters. In order to check the menu
parameters in the Status-/Options Menu, select "Status" respectively "Options" in the
Main Menu, open menu with
l and scroll through menu with t.
2.2 Rate, VTBI and Time Change Without Infusion
Interruption and Reset of Status Menu Data
• Press c when the pump is in the run display in order to switch to the Main Menu.
Select rate/VTBI/time with
t and press l in order to open the parameter.
•
Enter new value with q and confirm with k.
Reset Status Menu Data:
The parameters intermediate volume and -time can be reset when the pump is
infusing or when the pump is stopped.
•
Select “Status” in Main Menu with t and press l.
•
Highlight intermediate volume (in ml) or intermediate time (in h:min) with t and
open parameter with
l.
• Reset values by pressing u.
Both parameter total volume and -time, are displayed in the pump as "Total" with the
according unit and can be reset by starting a new therapy. A second way to reset the
parameters when the pump is in the Main Menu: Press
c, answer question if the last
therapy is to be used with
u and reset the values with u.
The type of the inserted syringe is displayed in menu item „Syringe“ and cannot be
changed once it has been confirmed at the beginning of the infusion. The drug info
states the drug name, the name of the drug list and its date of origin. The current
battery capacity in hours and minutes is displayed in the menu item “Battery Cap.” and
the current software version in menu item "Version". In-line pressure can also be read
in the Status menu in mmHg or Bar depending on the service settings.
19
Chapter 1
OPERATION