54
Activation of the function:
Open data lock in Options Menu with
l.
Select between level 1, 2 or 3 (if activated) with
l and r and confirm
with
k.
Enter code with
q and press k in order to activate data lock.
Changes to the protected values and the bolus function which are marked
with
y are only possible after entering the code. After 20 sec in the Main
Menu, Status Menu, Special Functions Menu and Options Menu the lock will be
activated again. If the wrong code is entered twice the pump will switch into
the last menu. If the wrong code is entered twice again the pump will go into
an audible alarm, a nurse call will go off and the yellow LED blinks. If a target
value was reached while data lock is active a new start of the pump is only
possible after entering the code.
In order to deactivate the function, select “Off” in the data lock, press
k, enter
the code and press
k again.
5.4 Bolus Rate
Open bolus rate in Options Menu with l.
Change bolus rate with q and confirm setting with k.
Note: Set bolus rate according to therapy requirements. Take care not to overdose!
Given a bolus rate of 1800 ml/h, e.g. 0,5 ml are reached within just one second.
5.5 KVO-Mode
After reaching a preselected VTBI/time, the pump can continue the infusion with a
predefined KVO-rate (see "Technical Data"). The duration of the KVO-infusion is set via
the service program.
Open KVO-Mode in Options Menu with l.
Answer Yes/No question with u, to activate KVO.
5.6 Contrast / Display Light / Keypad Light
Contrast as well as display- and keypad light can be adjusted individually according to
the lighting conditions.
Open contrast/display light/keypad light in Options Menu by pressing l.
Choose between 9 contrast- and display light levels with l or r and confirm
with
k. For use with light sensitive drugs the keypad- respectively syringe light
can be completely turned off.
Chapter 5
OPTIONS