Chapter 5
ALARMS
ALARMS
The Perfusor® Space is equipped with a audible and optical alarm signal.
6.1 Device Alarms
When a device alarm occurs the infusion is immediately stopped. Press O to switch
off the device. Then switch the device on again. In case of a repeated device alarm you
must disconnect from the patient, open the front door of the pump and take out the
disposable. The device needs to be handed to the service department.
6.2 Pre-Alarms and Operating Alarms
Pre-alarms:
Pre-alarms occur a few minutes (dependable on service settings, excluding OccluGuard
and pressure leap/drop pre-alarms) prior to operating alarms. During pre-alarms an au-
dible tone sounds, the yellow LED is constantly on and a staff call is activated (optio-
nal). The display message varies depending on the alarm reason. The signal tone and
the staff call are turned off with
c . Display and LED stay in pre-alarm until the ope-
rating alarm goes off. Pre-alarms don't lead to an interruption of the infusion.
Alarm-
type
Device
Alarm
yes flashes off device
alarm and
alarm code
(see service
program)
yes Press O and follow
the instruction on the
display.
Op
er
a
-
t
ing
A
-
lar
m
y
es
f
lashes
o
f
f
see al
arm
d
esc
ri
p
t
io
n
yes
Pre-
Alarm
yes off constant
on
see alarm
description
(de-)activate
via service
program
Press c to mute
alarm and turn off
staff call. Visible
alarm remains until
end.
Reminder
Alarm
yes off constant
on
see alarm
description
yes
Press k to mute
alarm, turn off staff
call and delete the
alarm text.
Alarm
Hint
no off
off
see alarm
description
no
Hint disappears
without confirmation.
Audible
signal
Red LED
Optical signal
Yellow LED Text
Staff call User confirmation
P
r
ess
k
t
o
ac
k
no
wled
g
e
t
he
aud
ib
le
alar
m,
alar
m
t
ext
and
st
af
f
c
all.
The
r
ed
LED
g
o
es
o
f
f
.
P
r
ess
c
t
o
mut
e
t
he
alar
m
f
o
r
2
minut
es.

Table of Contents