OPERATION
1.4 Syringe Change and New Therapy Start
Note: To avoid incorrect dosing, always disconnect the pump from the patient when
changing the syringe. Never leave the pump device unattended during syringe change.
Before inserting a new syringe check if the axial fixation is properly working.
Press sf to stop the infusion. The green LED will disappear.
Disconnect the pump from the patient.
Either the drive mechanism moves backwards into starting position when opening
syringe holder or the question "Perfom change?" must be first answered before the
drive moves backwards.
Open pump door, remove syringe and insert new syringe.
Note: In case the plunger head of the syringe is not released anymore by the claws
when performing a syringe change, the emergency release button needs to be pressed
to release the claws of the drive head. The emergency release button is placed on the
outside of the drive head. It can be released with a pointed pen. Then manually open
the claws and take out the syringe.
Close the syringe holder (Note: Piston brake must move forward!) and the pump
door and confirm the inserted syringe type with
k. Drive advances and grips
pressure plate of syringe.
Note: Do not block advancing drive unit with any objects. Piston brake must move
backwards into the syringe holder.
Prime pump if necessary with u then press d to continue.
Connect the patient to the pump and check set parameters using t.
Press sf to start infusion.
To start a new therapy after a syringe change:
Press c when pump is in the Main Menu.
Press d and continue to set new therapy parameters with q.
Press sf to start infusion.
Note: A new therapy can be started at any time during a stopped infusion. Press
c (repeatedly) when the pump is in the Main-, Status- or Options Menu and
proceed to follow instructions as described.
1.5 End of Infusion
Press sf to stop the infusion. The green LED disappears. Disconnect the
pump from the patient.
Open the syringe holder. Answer the question whether a syringe change should be
performed with
u. The drive moves backwards into the starting position.