36
The pump can be started now by pressing sf.
Starting Program Mode via Special Function Menu:
• Switch on pump with
o and wait until self-check is finished.
• Insert disposable.
• Go to Special Functions Menu and select Program Mode.
• Press
l to enter parameters and k to confirm.
• Adjust VTBI with
q.
After entering all desired parameters the pump can be started by pressing
sf.
In the upper part of the display the icon for “Program Mode” appears.
The screen shows the following:
Only the VTBI may be changed during an infusion that is running.
• Press
C to check upcoming Program Mode intervals in Main Menu.
It is possible to cancel one step of the running therapy. All following steps in
the programmed sequence persist.
• Go to Main Menu by pressing
C.
• Use
t to navigate through the Main Menu and select Current with l.
• For checking upcoming intervals press
C.
• Select "Program Parameters" with
l.
• Go through all interval steps with
r.
Note: The delivery of drugs can be stopped and started again in the Program
Mode at any time by pressing
sf. This will not have any effect on the settings
of the therapy.
Number of cycles is defined by VTBI. Take care to set the VTBI in the correct
relation to the volume of one Cycle. VTBI may needs to be adjusted after
changing the intervals.
The Main menu informs about the current interval. The configured parameters
can be checked by Program Parameter Menu in Main.
3.9 Intermittent Mode
The Intermittent Mode consists of 2 phases. This phases will be repeated.
•
Bolus phase: the configured bolus is active
•
Rate phase: time during the therapy in which the entered rate is active
The pump infuses the
predefined rate in the
predefined time for the current
step.
SPECIAL FUNCTIONS
Chapter 3