Changing the time interval after start:
If the user edits the time interval the therapy progression changes.
Press
C.
Use
t to select Interval and press l.
Change Interval by using
t and press k to confirm.
The pump automatically recalculates all other settings of the therapy.
3.10 Dose Over Time
Dose Over Time is used to administer a specific dose of antibiotics in a specific
time. Dose Over Time is an own therapy and cannot be used in combination
with another therapy. It can only be activated via the Drug List Manager. It can
be used for standard infusion.
The active Dose Over time function is always symbolised with a characteristical
symbol in the Display ( ).
Note: Dose Over Time should only be performed by experienced users being
familiar with the principles of the Dose Over Time function and properly trained
in using the present device.
The infusion rate in Dose Over Time can not be changed. This parameter is a
result of the total dose and the infusion time setting. Directly, after the Drug
selection, the infusion time and the total dose intended to be infused have to be
set. If the drug library contains default values for these parameters, the default
values are used as preset values.
If changes are necessary during infusion, the delivery can be controlled by
changing the time. The pump calculates the new rate by using the remaining
total dose and the remaining time. In the Main Menu total dose, time and VTBI
can be changed, also during RUN-Mode. Other parameters (dose rate, basal rate,
concentration, patient weight and patient height) cannot be changed.
Note: The KVO function and Bolus function are disabled during Dose Over Time.
Note: The feature Dose Over Time always requires the usage of dosing units (i.e.,
mg or mg/kg patient weight).
Before using Dose Over Time contact your local B. Braun representative!
Starting Dose Over Time via Drug Library:
Note: Dose Over Time settings have been configured in the Drug List Manager
before and have been uploaded into the pump.
Switch on pump with
o and wait until self-check is finished.
SPECIAL FUNCTIONS
Chapter 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
BrandBraun
ModelPerfusor Space
CategoryHealth Care Products
LanguageEnglish