20
CALIBRATION “C” NUMBERS OR CALDAC (see Calibration of the Analog Output)
The display will show all four (4) calibration or “C” numbers.
Each number may be up to three (3) digits in length. By
recording these numbers you will be able to return the
748P to its present calibration settings without using test
weights simply by entering the four (4) “C” numbers. Note
that your scale is used in a commercial application for
buying or selling, you may not use the “C” numbers. The
“C” numbers may be reviewed during normal operation by
pressing the TEST key.
CALIBRATION/SETUP COMPLETED
The indicator will reset, ready for operation. NOTE: The indicator can be turned off anytime during
the calibration and setup procedure with all previous data retained.
FINE SPAN ADJUSTMENT
With power applied and displaying weight, depress and hold the calibration switch. Then press 0 to
decrease span or 1 to increase span. Release the switch to return to normal operation.
C4=
C3=
C1=
C2=