The following are required for disinfection/clean-
ing:
ü
Non-foaming disinfectant/cleaning agent that
is compatible with the materials.
ü
Unit care system, e.g. OroCup
❯
To pre-clean, suck up 2 litres of water with the
care system.
❯
Aspirate the disinfection/cleaning agent with
the care system.
13.3 Once or twice a week before
the midday break
Under harsher conditions (e.g. hard water
or frequent use of prophy powders) 1x
daily before the midday break
The following are required for cleaning:
ü
Special non-foaming detergent for suction
units that is compatible with the materials.
ü
Unit care system, e.g. OroCup
❯
To pre-clean, suck up 2 litres of water with the
care system.
❯
Aspirate the cleaning agent with the care sys-
tem.
❯
Rinse with ca. 2 l water after the application
time.
14 Replace the amalgam col-
lector vessel
NOTICE
Risk of contamination if the amalgam
collector vessel is reused since the
collector vessel is not water-tight.
❯
Do not use the collecting container
more than once (disposable item).
Wear protective equipment to avoid any
risk of infection (e.g. liquid-tight protective
gloves, protective goggles, face mask).
We strongly recommend that the amal-
gam collecting container should only be
changed in the morning before the start
of work. This will prevent fluid from drip-
ping out of the drum while it is being
changed.
❯
Disconnect all power from the unit.
❯
Remove the full amalgam collecting container
and from the device.
❯
Pour disinfectant for suction units (e. g. Orotol
plus, 30 ml) into the full amalgam collecting
container.
❯
Close and secure the full amalgam collecting
container using the cap. Observe the markings
on the cap and on the collecting container.
❯
Place the securely closed amalgam collecting
container into its original packaging and seal.
❯
Insert a new amalgam collecting container in
the unit and clamp it in position.
Only use original amalgam collecting con-
tainer.
❯
Switch on the power supply. The unit is ready
for operation again.