Internal Variables
78 ID_COMPENSATED_EXTRA_AD2
This input is available only on “F” models. This is the
compensated external analog input value, after linearization
has been applied to the raw signal. A value of 6400 is 0% and
59200 is 100%.
85 ID_AD_EXTRA1_TOGGLE
This variable is used to adjust the level that the ER5000
interprets variable #77 (ID_COMPENSATED_EXTRA_AD1) as
a Logical 0 or a Logical 1.
86 ID_AD_EXTRA2_TOGGLE
This variable is used to adjust the level that the ER5000
interprets variable #78 (ID_COMPENSATED_EXTRA_AD2) as
a Logical 0 or a Logical 1.
ER5000 Pressure Profile Variables
45 ID_PROFILE_LOOP_COUNT
This variable displays the count of times through a Prole
loop. The counter starts at the number of loops entered in
the Number of Loops (0=>forever)eld of the Prole Builder when
the Loop command segment is added to the Prole and
decrements by one at each loop of the count.
69 ID_PROFILE_TYPE
This is the type of the prole step currently active.
0 => STOP
1 => RAMP
2 => STEP
3 => DWELL
4 => DIGITAL OUTPUT
5 => DIGITAL INPUT
6 => LOOP
7 => CHANGE VARIABLE
8 => Delta
9 => If
10 => GoTo
11 => Soak

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Tescom ER5000 Series and is the answer not in the manual?

Emerson Tescom ER5000 Series Specifications

General IconGeneral
BrandEmerson
ModelTescom ER5000 Series
CategoryController
LanguageEnglish

Related product manuals