A-2 ApexPro™ 2001989-351A
Technical specifications
Below are the technical specifications for the ApexPro telemetry system.
NOTE
Due to continual product innovation, GE designs and specifications are subject to
change without notice. Contact your sales/service representative for the most
current information.
ApexPro and ApexPro CH transmitter
Performance specifications
Power requirements
Alarms and controls
Transmission
ECG
Battery type: ANSI/NEDA 15A, 1.5 V AA alkaline (2 required)
Battery life: ApexPro model: 40 hours typical
ApexPro CH model: 120 hours typical
Polarity: Electronic reverse polarity protection
Battery integrity: Transmitted and indicated via LED
Lead fail indication: Transmitted and indicated via LED
Channel spacing: 25 kHz
Frequency stability: ± 0.0001% of assigned channel frequency
Bit rate: 10 kb/sec
Antenna: Formed by leadwire shield
Multi-channel (5- or 6-
leadwire) configuration:
I, II, III, Va, Vb, aVR, aVL, aVF
Leads analyzed
simultaneously:
Four (I, II, III and V)
Single-channel (3-leadwire)
configuration:
I, II or III, programmable
Heart rate detection range: 30 to 300 beats/minute