Maintenance
2001989-351A ApexPro™ 4-5
Verify labels
Follow these steps to be sure that the TTX number shown on the transmitter is the
same as the programmed TTX number:
1. Verify that the Switch Label is present and securely attached to the front of the
case.
2. Determine the programmed TTX number. For more information, see View the
TTX number on page 3-19.
3. Verify that the data recorded on the TTX label on the back of the transmitter is
accurate for the transmitter.
CAUTION
Make sure the TTX numbers match. Failure to do so may result in
monitoring the wrong patient.
Cleaning
All equipment should be cleaned on a regular basis. Comply with the policies of your
institution’s infection control unit and/or biomed department. The decision to
disinfect or sterilize must be made per your institution’s requirements with an
awareness of the effect on the integrity of the transmitter and leadwire.
WARNING
Disconnect AC-powered equipment from the power line before
cleaning or disinfecting its surface. Turn off the power to battery-
powered equipment before cleaning or disinfecting its surface.
CAUTION
Never immerse devices, cables, or leadwires in any liquid.
CAUTION
Do not pour or spray any liquid directly on cables or leadwires or
permit fluid to seep into connections or openings.
CAUTION
Never use conductive solutions, solutions that contain chlorides,
wax, or wax compounds to clean devices, cables or leadwires.