EasyManuals Logo
Home>GE>Medical Equipment>ApexPro TT

GE ApexPro TT User Manual

GE ApexPro TT
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #183 background imageLoading...
Page #183 background image
Electromagnetic compatibility
2001989-351A ApexPro™ B-5
Recommended separation distances
The following table provides the recommended separation distances (in meters)
between portable and mobile RF communications equipment and the ApexPro
Telemetry Transmitter.
The ApexPro and CARESCAPE Telemetry Transmitter is intended for use in the
electromagnetic environment on which radiated RF disturbances are controlled. The
customer or the user of the ApexPro Telemetry Transmitter can help prevent
electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the ApexPro and
CARESCAPE Telemetry Transmitter as recommended below, according to the
maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance [d] in meters (m) can be estimated using the equitation applicable
to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all instances. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance in meters (m) according to frequency of transmitter
Rated maximum output
power of transmitter in watts
150 kHz to 80 MHz
a
d = 11.6
80 MHz to 800 MHz
a
d = 11.6
800 MHz to 2.5 GHz
a
d = 23.3
0.01 1.16 1.16 2.33
0.1 3.67 3.67 7.37
1 11.6 11.6 23.3
10 36.7 36.7 73.7
100 116 116 233
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro TT and is the answer not in the manual?

GE ApexPro TT Specifications

General IconGeneral
BrandGE
ModelApexPro TT
CategoryMedical Equipment
LanguageEnglish

Related product manuals