A-8 ApexPro™ 2001989-351A
Technical specifications
Apex oximeter
Performance specifications
Physical specifications
Certification
UL 60601-1 CE marking for the 93/42/EEC Medical Device Directive.
Battery Type: ANSI/NEDA 15A, 1.5V AA alkaline (2 required)
Battery Life: 30 hrs. typical with display on continuously; 60 hrs. typical
with display off
Polarity: Electronic reverse polarity protection
Water resistant with sensor
in place:
IEC 60529 IPX3 rating (spray and wipe only)
Controls and Indicators
Power on/off: Button switch
Display on/off: Button switch
LED indicator: Numerical SpO2 value; numerical HR value; signal quality
indicator
SpO2 sensors: Full line of Nonin qualified reusable and disposable sensors
Processing
Saturation range: 0 to 100%
Saturation accuracy: 70 to 100% ± 3 digits (± 1 S.D.)
Pulse rate range: 18 to 300 BPM
Pulse rate accuracy: ± 3 % or 1 BPM, whichever is greater
CIC Pro center
Messages: Check probe, Low signal quality, Probe off patient
Numerical display: SpO2 (% value), HR (bpm)
Alarms: SpO2 high/low, HR high/low
Height: 12.7 cm (5.0 in)
Depth: 2.0 cm (0.8 in)
Width: 6.9 cm (2.7 in)
Weight: 113.4 g (4.0 oz) Weight excludes batteries and SpO2 probe