EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P5

GE LOGIQ P5 Service Manual

GE LOGIQ P5
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 7-3 - Screen Captures 7-5
7-3-1 Check and Record the P2 Key Function (cont’d)
5.) In the “Physical Print Buttons”box, record the followings.
* Format
* Image Frames
* Capture Area
* Compression
6.) In the “Printflow View”, record the print flow associated with the Print2 Button.
7-3-2 Setting the P2 Key to Screen Capture
If the P2 Key is not set to screen capture:
1.) While on the Connectivity screen, with the Buttons tab displayed, go to the drop down selection
menu in the
Destinations section.
2.) Ensure that Capture Area is set to Whole Screen, secondary Capture and None Image
Compression.
3.) The P2 Key should now be set up for whole screen capture, sending the screens to the image
buffer (clipboard).

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5
CategoryMedical Equipment
LanguageEnglish

Related product manuals