DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 2-3 - Facility Needs 2-9
2-3-4 Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
LINEN SUPPLY
EXAMINATION
TABLE
FILM VIEWER
COUNTER TOP
FILE
CABINET
SECRETARYS OR
DOCTOR’S DESK
FILM
PROCESSING
ROOM,
OVERHEAD
LIGHTS DIMMER
DEDICATED
POWER
RECEPTACLE
EXTERNAL
PERIPHERALS
SINK
COUNTER
TOP
FILM VIEWER
FILM SUPPLIES
SINK
LINEN SUPPLY
PROBES/SUPPLIES
24 IN.
(61 CM)
EXAMINARION
TABLE
76 IN.
(193 CM)
DOOR
30 IN.
(76 CM)
An 8 by 10 foot Minimal Floor Plan
DEDICATED POWER
OUTLETS
DEDICATED ANALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
A 14 by 17 foot Recommended Floor Plan
GE CABINET
FOT SOFTWARE
AND MANUALS
Scale : Each square equals one square foot
DOOR
42 IN.
(107 CM)
DEDICATED ALALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
SUCTION LINE
EMERGECY
OXYGEN
24 IN.
(61 CM)
76 IN.
(193 CM)
PATIENT
TOILET
FACILITY
FOOT
SW
FOOT
SW
LOGIQ 5
CONSOLE
CONSOLE
18 IN.
(46 CM)
STOOL
STOOL

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5
CategoryMedical Equipment
LanguageEnglish

Related product manuals