GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-109
8-2-48 ACPC Assy
8-2-48-1 Tools
• Common pilIips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-48-2 Preparations
• Shut down the system and switch off the main breaker.
8-2-48-3 Removal procedure
1.) Remove the AC Input Box Assy. Refer to the 8-2-49 "Transbox Assy 110V/220V" on page 8-109.
2.) Disconnect 5 connectors.
3.) Unscrew 5 screws to remove the ACPC Assy.
Figure 8-146 ACPC in Transbox
Figure 8-147 Screws of ACPC assy