DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-82 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-33 SYSCON PM Assy/SYSCON CM Assy
8-2-33-1 Tools
• Common pilIips screwdrivers
8-2-33-2 Preparations
• Shut down the system and switch off the main breaker.
8-2-33-3 Removal procedure
1.) Remove the Side Right Cover. Refer to the 8-2-16 "Side Right Cover" on page 8-52.
2.) Remove the EMI Cover R. Refer to the 8-2-27 "EMI Cover R" on page 8-69.
3.) Remove the SYSCONPM(SYSCONCM) assy in slot #3 in the nest box.
4.) Eject one ASIR ROM from old SYSCONML with regular screw driver.
Figure 8-116 PWAs in Nest Box