GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-110 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-48-3 Removal procedure (cont’d)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-48-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-49 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-48. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-12
ACPC Assy set Function Validation Procedure
Section 10-5-2-2
Peripheral/Option Checks
Section 10-5-5
Physical Inspection
Section 10-6-2
GEMS Leakage Current Limits
Section 10-6-3
Outlet Test - Wiring Arrangement - USA &
Canada
Section 10-6-4
Grounding Continuity
Section 10-6-5
Chassis Leakage Current Test
Section 10-6-6
Isolated Patient Lead (Source) Leakage–Lead
to Ground
Section 10-6-7
Isolated Patient Lead (Source) Leakage–Lead
to Lead
Section 10-6-9
Probe Leakage Current Test

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P5
CategoryMedical Equipment
LanguageEnglish

Related product manuals