6 FC1: Legally Relevant Functionalities
Page 48 of 61
Use in Legal Metrology
enCore FC1
6 FC1: Legally Relevant Functionalities
6.1 Basic System
The main task of the Basic System is the I/O board interfacing and the digital
protocol interfacing for intelligent measurement devices like USMs or PGCs.
That means that the Basic System collects, monitors, and evaluates all legal-
ly relevant input data coming in via process boards and protocol ports. It is
able to perform adjustments for pressure and temperature measurements as
well as a geometry correction for ultrasonic gas meters, if parameterized.
From the fiscal point of view, the Basic System provides all measurement
data and status information needed for the volume conversion for the
Gas
Quality
and Flow Conversion AFBs.
In detail:
Data source Legally relevant data processed by the Basic
System
Gas meter Meter progress
Flow rate
Flow direction information (if applicable)
Status information
Pressure transmitter Measurement, adjusted measurement
Status, error information
Temperature
transmitter
Measurement, adjusted measurement
Status, error information
Gas quality
measurement device
Measurements
Status, error information
Table 13: Legally relevant measurement data

Table of Contents

Other manuals for Honeywell enCore FC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell enCore FC Series and is the answer not in the manual?

Honeywell enCore FC Series Specifications

General IconGeneral
BrandHoneywell
ModelenCore FC Series
CategoryDesktop
LanguageEnglish

Related product manuals