EasyManuals Logo
Home>Honeywell>Security Sensors>SPM Flex

Honeywell SPM Flex User Manual

Honeywell SPM Flex
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
34
SPM Flex Gas Detector
Optional sampling wand
The sampling wand is connected to the inlet port and used to detect toxic gas at
specific locations� Do not use the sampling wand with these gases:
diisocyanates
hydrazines
hydrogen peroxide
ozone
sulfuric acid
Do not use the sampling wand in highly humid conditions with these gases:
boron trifluoride
hydrogen bromide
hydrogen chloride
hydrogen fluoride
nitric acid
An inlet filter is installed in the grip to prevent debris from entering the detector
All gas-wet surfaces are either coated with or made of materials selected to be
compatible with sticky corrosive gases� Use an appropriate filter for the gas type
being monitored� Refer to “Detectable Gases” on page 62 for appropriate
filters and maximum tubing lengths� The sampling wand is supplied with the
mounting bracket and required hardware� The wand is connected to the detector
with the supplied coiled tube� The wand and the coiled tube are connected with
push fittings� To ensure a proper connection, simultaneously push and twist the
tubing into the fitting until it is fully seated, about one inch (2�5 cm)�
Use only Honeywell approved parts with this detector (see “Accessories and
Parts” on page 65)�
CAUTION
!

Other manuals for Honeywell SPM Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell SPM Flex and is the answer not in the manual?

Honeywell SPM Flex Specifications

General IconGeneral
BrandHoneywell
ModelSPM Flex
CategorySecurity Sensors
LanguageEnglish

Related product manuals