42
SPM Flex Gas Detector
Flow characterization
Flow characterization is an optional function� When utilized, it allows a flow
system to be optimized with a certain type of Chemcassette cartridge� This
allows the SPM Flex detector to quickly reach the correct flow rate�
Open/close gate
The gate can be opened and closed manually (from the Maintenance menu)� For
intermittent use, this allows Chemcassette cartridges to be conveniently removed
and stored�
Update program
Through the
Update program
option, the new firmware can be loaded via the
USB port�
Factory service mode
For Honeywell Analytics service personnel only�
Setup menu
Many of the functions of the SPM Flex detector can be configured from the Setup
menu�
General
The backlight intensity and timeout period are chosen from this menu as
well� The power-up options are chosen from this menu� The detector can be
programmed to be in monitoring mode or out of monitoring mode when powered
up� Out of monitoring mode is the default state� It can also be programmed to
power up in the state it was in when it was powered down� Finally, if languages
other than English are loaded in the detector’s software, the language can be
chosen from the General menu�
The detector IDs (“unit IDs”), either short (up to 20 characters) or long (up to
35 characters) and an idle timeout period of from 1 to 60 minutes (20 minutes
default) can also be entered from the General menu� An idle timeout period of 0
minutes will disable the idle timeout function�