63
SPM Flex Gas Detector
Specifications
Description Value
Detection Technique Chemcassette-cartridge-based with advanced self monitoring optics design
Dimensions see “Dimensions” on page 28
Weight 9.1 lbs. (4.1 kg)
Operating Temperature 0C to 40C for most gases / applications
Operating Humidity
0-100% RH for detector (Sample RH limited per tape/calibration). Sample line requires additional hardware to remove moisture in
high RH conditions where condensing may occur. The sample must be non-condensing. Dry conditions may require humidification.
Power input (from power adaptor/charger or direct-wired) 24 VDC ±10%, 3.75 amps
Power consumption ~1.9 A at 24 VDC ±10% (including battery-charging current)
Power adaptor/
charger
Manufacturer CUI Inc.
Model SD190-24-U-ST
Input 90~264 VAC, 1.2 A, 47-63 Hz
Output 24 VDC, 3.75 A max
CCN QQGQ (file #E210311)
Mark of conformity UL listed
Environmental use Indoor only
Communications
Relays: Alarm 1, Alarm 2, Fault (user configurable for normally open/closed)
4-20mA
Ethernet (with Modbus TCP/IP and web server)
USB port (for flash drive configuration/data transfer)
Flow System
Automatic flow control with bypass system, 250 or 500 cc/min at tape, higher flow at inlet to reduce sample time (internal
bypass system); sample up to 100 ft
Local Alarms/Status
Visual: LEDs for alarm, normal condition and fault
Audible: User selectable: Off, Low ~75 dB at 1 m, Medium ~85 dB at 1 m, High >90 dB at 1 m
Interface 4 large buttons, 3.5” Color LCD TFT display, web server
Data Logging Rolling 3 months (15 sec. with no gas reading, 1 sec. when reading gas), Event history (1500 events – approx. 1 year)
Mounting
screws
concrete
5/16 in x 2 in vibration-resistant stud anchor for concrete (McMaster-Carr 94475A185 or equivalent), add 0.25 in. to length
when mounting bracket with sun shield
wood
5/16 in. x 2 in. flange head lag screw for wood (McMaster-Carr 95526A375 or equivalent), add 0.25 in. to length when
mounting bracket with sun shield
Battery type Lithium ion
Battery life Approximately 70% of original capacity after 300 full charge/discharge cycles; 6+ hours under typical operating conditions
Maximum inlet/outlet pressure differential The overall maximum load on the pump between the inlet and the exhaust should not exceed 10 inches H
2
O
Communication connector, optional communications cable 60 V, 5 A maximum
Relays 250 V, 6 A maximum
Wire gauges
Minimum 20
Maximum 14
USB 2.0 or later
Indoor/outdoor use? both (the power adaptor/charger is indoor only)
Operating
altitudes
-1,000 to 3,000 ft. above sea level standard
above 3,000 ft. to 6,000 ft. above sea level requires factory adjustment, contact Honeywell Analytics
Ingress Protection rating IP65
External switch or circuit breaker requirement (description & location) meet or exceed all local codes and regulations
Ventilation requirements mount with no obstructions within 4 in. (10 cm) of either side or within 2 in. (5 cm) above and below the detector
4-20 mA output
defaults and
ranges
Inhibit 2 mA, programmable from 1.5-3.5 mA in 0.5 mA increments
Maintenance 3 mA, programmable from 1.5-3.5 in 0.5 increments
Instrument fault 1 mA or less, not programmable (will be driven under 1 mA)
Over-scale 21.5 mA, programmable 21-22 mA
4-20 mA configurations sink, source, isolated
Storage
conditions
Detector 32°F to 104°F (0°C to 40°C), 0-100% RH non-condensing
Chemcassette cartridges Refer to the label on the Chemcassette cartridge for storage conditions
Certifications
Detector
UL 61010-1, 3rd Edition, 2012-05 (ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL, AND LABORATORY USE -
Part 1: General Requirements
CAN/CSA-C22.2 No. 61010-1, 3rd Edition, 2012-05, (ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL, AND
LABORATORY USE - Part 1: General Requirements)
IEC 61010-1:2010, 3rd Edition
Global radio transmission approvals
Battery
UL/cUL Recognition to UL 2054 + 60950-1
IEC 62133 1st Edition CB Certification
UN Test Report to UN 38.3
Self-declared European CE Mark on detector for: EMC, LVD, ROHS, WEEE