EasyManuals Logo
Home>Honeywell>Security Sensors>SPM Flex

Honeywell SPM Flex User Manual

Honeywell SPM Flex
77 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
60
SPM Flex Gas Detector
Checking for system leaks
If the system uses the sampling wand or sample tubing, verify that the tubing is
correctly seating in the detector’s push fitting as shown in this figure�
Grey locking ring
Tubing
stop
Correct depth
(fully inserted)
TubingIncorrect depth
(not fully inserted)
O-ring (cross section)
Figure 38. Properly seating tubing in push fittings
In the Review menu, navigate to the Additional Status option� Plug the inlet
tube� If the system is not leaking, the detector will display a fault after about 30
seconds� This may take longer if the sampling wand is attached and possibly as
long as 2 minutes if the test gas must travel through a lengthy sample tube�
If a fault is not issued, observe the flow reading on the detector’s display� If it
stabilizes at more than the maximum rate shown in the “Sample Analysis Flow
Rate” column of “Detectable Gases” on page 62, there is likely a system leak�
Contact Honeywell Analytics’ Service Department�

Other manuals for Honeywell SPM Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell SPM Flex and is the answer not in the manual?

Honeywell SPM Flex Specifications

General IconGeneral
BrandHoneywell
ModelSPM Flex
CategorySecurity Sensors
LanguageEnglish

Related product manuals