D-0120578-C – 2019/11
Eclipse Instructions for Use - US
Page 34
Original curves Merged/summed curve
New Fmp and Residual noise values are calculated for the merged/summed curve, however the Fmp and
Residual noise curves will not be available.
A “M” is added to the intensity handle to indicate this is a Merged curve.
3.1.47 Adding curves
Curves of the same intensity, stimulus type and ear can be added together to generate a third curve, based
on the total average of the individual sweeps of the two selected curves.
1. Select one curve by doubling clicking on the curve’s handle.
2. Right click on the second curve and select Add this curve. The added curve will be indicated by an
‘a’ following the intensity on the handle (e.g., 60a)
Original curves 60a is the sum of the two other curves
The added curve can be removed/deleted by selecting Delete when right clicking on the curve.
New Fmp and Residual noise values are calculated for the added curve, however the Fmp and Residual
noise curves will not be available.
3.1.48 Differential Ipsi minus contra (Ipsi-Contra) curve
Right click on the curve’s handle to show the differential Ipsi minus Contra
curve.

Table of Contents

Other manuals for Interacoustics Eclipse

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Eclipse and is the answer not in the manual?

Interacoustics Eclipse Specifications

General IconGeneral
BrandInteracoustics
ModelEclipse
CategoryMedical Equipment
LanguageEnglish

Related product manuals