12-6 SCPI Tables
Table 12-4
SENSe command summary (refer to Tables 2-3, 3-2, and 4-2)
Command Description
Default
paramete
r
SCPI
:SENSe[1] SENSe subsystem for channel #1 (battery channel)
:FUNCtion <name> Select measurement function (“VOLTage,” “CURRent,”
“D
VMeter”, “PCURrent,” or “LINTegration”).
VOLT
:FUNCtion? Query measurement function.
:NPLCycles <n> Specify integration rate (in line cycles) for voltage, current,
and D
VM measurements (0.01 to 10).
1 ✓
:NPLCycles? Query integration rate. ✓
:AVERage <NRf> Specify the average count for voltage, current, and DVM
measurements (1
to 10).
1
:AVERage? Query average count.
:CURRent [:DC] Path to configure the current measurement function: ✓
:RANGe Current measurement range: ✓
[:UPPer] <NRf> Specify expected current measurement which will
se
lect 5mA or 5A amp range for the Models 2306
and 2306-VS, or 500mA or 5A for the 2306-PJ.
5.0 ✓
[:UPPer]? Query current measurement range. ✓
:AUTO <b> Enable or disable auto range. OFF ✓
:AUTO? Query state of auto range.
:PCURrent Path to configure the pulse-current measurement function
(see Section
3):
:AVERage <NRf> Specify average count (dependent on SYNC:STAT):
pulse
-current measurements: 1 to 100 (sync ON)
pulse current digitization: 1 to 5000 (sync OFF)
1
:AVERage? Query average count (dependent on SYNC:STAT).
:MODE <name> Select pulse-current measurement mode (HIGH, LOW,
or A
VERage).
HIGH
:MODE? Query pulse-current measurement mode.
:TIME Path to set pulse-current integration times:
:AUTO Power supply sets integration times — accounts for
internal trigger delay
but not user delay (:DEL).
:HIGH <NRf> Specify integration time (in sec) for high pulse
measurements
(33.33E-06 to 0.8333) — accounts
neither for internal trigger delay nor user delay.
3.333E-05