61 PBS/PBJ-BA-e-1718
16.9 Formulation mode
This mode is used for convenient weighing of individual recipe components. Weights
of each component are displayed and saved after each pressing the button. The
weights of those component are sent with the RS-232C or DATA I/O interface, and
the display is automatically zeroed to weigh the next component.
When all components are weighed, their weights are added up and the total weight is
displayed. This value output is edited by the button.
Select the menu element 51 in the menu overview.
• 1. If the recipe mode is activated, the display is in the recipe stand-by mode
until it is started. The display shows the Add-On symbol, memory symbol and
stand-by mode symbol. Place the container (if used) and press the
button to tare it. Please remember that taring with the button will not
be accepted after previous pressing the button (as in the step 2). Taring
is possible again after previous pressing the button (as in the step 5).
• 2. Press the button. If a peripheral device is connected,
„--------- FORMULATION MODE ---------” will be emitted.
• 3. Place the first component and then press the button. The weight value
will be output as „CMP001”. After that, the display will be automatically
zeroed.
• 4. Step 3 is to be repeated for all components to be weighed.
• 5. Then press the button. The total weight will be displayed and output to
peripheral devices with the „TOTAL=” unit.
• 6. Remove everything from the balance plate. The next recipe is started from
the step 1.