PBS/PBJ-BA-e-1718 62
16.10 Automatic saving and zeroing
This function is used for weighing large number of single samples. If the function is
activated, the automatic saving and zeroing symbol is illuminated.
Select the menu element 52 in the menu overview.
• Place the container to be weighed on the balance and press the
button when the automatic saving and zeroing function is in the stand-by
mode. (The automatic saving and zeroing symbol and mode
symbol are illuminated.)
The balance will be zeroed.
• Press the button. The stand-by symbol disappears and measurements
within the automatic saving and zeroing function can be started.
• Place the first sample on the scale pan. Each time when the stabilisation
symbol is illuminated and the value corresponding to five times the zero range
or higher is illuminated or when the button is pressed, the illuminating
value output and balance zeroing is carried out.
• Weighing process of the next sample is carried out without the necessity to
press the button.
• Press the button. The balance returns to the stand-by mode of the
automatic saving and zeroing function, the total weight placed on the scale
pan without the packaging weight is displayed. To print this value, press the
button.
Notes:
• If the stabilisation symbol is illuminated and the display value lies within the
zero range, zeroing is carried out automatically.
• Pressing the button when the display value lies below five times the zero
range is followed by zeroing after data output. (Manual loading)
• Pressing the button when the automatic saving and zeroing function is in
stand-by mode is also followed by setting mains supply in stand-by mode.