5.2 Various problems and countermeasures
111
Chapter 5
5.2.3 AeroDR Interface Unit2
Status Error description Corrective actions
Power LED (green) does not
light.
AC power not supplied.
Make sure the power cable is connected cor-
rectly.
Feeding1/2 LED (blue) does not
light.
Feeding1/2 LED (blue) does not light even
when connecting the DR Detector.
Make sure that the I/F Cable is connected
correctly
.
Communications are down
between devices, and power
LED (green) is o󰀨.
The power switch is not on. Or, the power
cable is not connected.
Make sure the power switch of the AeroDR
Interface Unit2 is on. Or, make sure the power
cable is connected correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller, and
feeding1/2 LEDs (blue) are o󰀨.
The I/F Cable is not connected.
Make sure that the I/F Cable is connected to
the DR Detector properly.
Battery is not charging while the
DR Detector is used with wired
connection (the battery level on
the battery level display of the
image processing controller does
not increase), and feeding1/2
LEDs (blue) are o󰀨.
5.2.4 Detector Interface Unit
Status Error description Corrective actions
LED(green/blue) does not light. AC power not supplied.
Make sure that the Ethernet cable is
connected to the Power Supply Unit properly.
Or, make sure that the DI Unit AC adapter is
connected correctly.
LED (blue) does not light.
The I/F cable is not connected to the DR
Detector.
Make sure that the DR Detector is connected
to the I/F cable correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image pro-
cessing controller.
The I/F cable is not connected to the DR
Detector. Or, the Ethernet cable is not
connected to the Power Supply Unit or the
hub.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta AeroDR System and is the answer not in the manual?

Konica Minolta AeroDR System Specifications

General IconGeneral
BrandKonica Minolta
ModelAeroDR System
CategoryMedical Equipment
LanguageEnglish

Related product manuals