5.2 Various problems and countermeasures
113
Chapter 5
5.2.8 AeroDR Generator Interface Unit2
Status Error description Corrective actions
LED (green) does not light. AC power not supplied.
Make sure that the Ethernet cable is
connected correctly.
Make sure the power switch of the AeroDR
Interface Unit is on.
If the AeroDR Generator Interface Unit2 has
a power switch, make sure the power switch
is on.
LED (green) does not change
from ashing to a solid light.
There is a communications error.
Contact Konica Minolta technical
representatives.
5.2.9 Generator Interface Unit 3
Status Error description Corrective actions
LED (blue) does not light.
AC power is not supplied.
Make sure that the Ethernet cable is
connected correctly.
Make sure the power switch of the Power
Supply Unit is on.
If GIU3 AC Adapter is used, make sure the
power cable is connected correctly.
There is a communications error.
Contact Konica Minolta technical
representatives.
5.2.10 AeroDR Battery Charger
Status Error description Corrective actions
Status LED (green) does not
light.
AC power not supplied.
Make sure that the power cable is connected
correctly.
Error LED (orange) is on. -
Contact Konica Minolta technical
representatives.
When the DR Detector
*1
is
inserted, the error LED (orange)
lights.
DR Detector was inserted the wrong way.
Check the insertion orientation of the
DR Detector.
The DR Detector is not inserted correctly.
Make sure that the DR Detector is inserted
correctly.
Although the DR Detector
*1
is
inserted, charging does not start,
and the status LED (green) is out.
AC power not supplied.
Make sure that the power cable is connected
correctly.
Although the DR Detector
*1
is
inserted, it cannot be registered, and
the status LED (green) is ashing.
The Ethernet cable is disconnected.
Make sure that the Ethernet cable is
connected correctly.
*
1
The AeroDR
1012
HQ and AeroDR
3
1012
HQ cannot be inserted into the AeroDR Battery Charger.