119
6.1 Maintenance and inspection items
Chapter 6
6.1.4 Consumables
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Refer to each device’s manual for information
about periodic replacement parts and consumables
for the image processing controller, etc.
In particular, continued use of the battery may re-
sult in degradation and wear, and it may no longer
exhibit proper functioning capabilities. For extend-
ed, safe use, it is necessary to replace parts which
have become worn or degraded.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
6.1.5 AeroDR Grip sheet a󰀩x-
ing and replacement
When a󰀩xing the AeroDR Grip sheet to the DR Detector,
follow the procedure below.
Application locations
AeroDR Grip sheet
AeroDR Grip sheet
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When using a tool to a󰀩x the AeroDR Grip sheet or
perform replacement, be careful not to get injured
by a tip or edge of the tool.
For AeroDR Grip sheet affixing or replacement, it
is recommended to ask Konica Minolta technical
representative.
The AeroDR 1012HQ, AeroDR 3 1417HD, AeroDR
3 1717HD and AeroDR 3 1012HQ do not use the
AeroDR Grip sheet.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Remove the label from the AeroDR Grip
sheet a󰀩xing position.
Insert a precision flat-blade screwdriver, etc. into
the step at the AeroDR Grip sheet a󰀩xing position
to slightly turn the label over, and then remove it
completely.
Label on the AeroDR Grip
sheet a󰀩xing position
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Be careful not to scratch the DR Detector.
Be sure to peel o󰀨 the label or used AeroDR Grip
sheet because it may come o󰀨.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Remove dust and contamination from the
AeroDR Grip sheet a󰀩xing position.
AeroDR Grip sheet a󰀩xing position
HINT
•••••••••••••••••••••••••••••••••••••
After removing the label or used AeroDR Grip
sheet, clean o󰀨 any adhesive residue on the sur-
face using a soft cloth moistened with anhydrous
ethanol.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta AeroDR System and is the answer not in the manual?

Konica Minolta AeroDR System Specifications

General IconGeneral
BrandKonica Minolta
ModelAeroDR System
CategoryMedical Equipment
LanguageEnglish

Related product manuals