5.2 Various problems and countermeasures
112
5.2.5 Detector Interface Unit 2
Status Error description Corrective actions
LED(green/blue) does not light. AC power not supplied.
Make sure that the Ethernet cable is
connected to the Power Supply Unit properly.
Or, make sure that the DI Unit AC adapter is
connected correctly.
LED (blue) does not light.
The I/F cable is not connected to the DR
Detector.
Make sure that the DR Detector is connected
to the I/F cable correctly.
The DR Detector is being used
with wired connection, but it is
not recognized on the image
processing controller.
The I/F cable is not connected to the DR
Detector. Or, the Ethernet cable is not
connected to the Power Supply Unit.
Make sure that the DR Detector is connected
to the I/F cable correctly. Or, make sure that
the Ethernet cable is connected correctly.
Serial exposure does not start.
The Ethernet cable for relaying serial exposure
signals is not connected to the Sync port.
Make sure that the Ethernet cable for relying
serial exposure is connected to the Sync port.
5.2.6 Power Supply Unit
Status Error description Corrective actions
LED does not light. AC power not supplied.
Make sure the power cable is connected cor-
rectly.
The power switch cannot be
turned on.
There is an error in the power system.
Contact Konica Minolta technical
representatives.
The DR Detector is being used
with wired connection, but it is
not recognized on the image pro-
cessing controller.
The Ethernet cable is disconnected.
Make sure that the Ethernet cable is
connected correctly.
5.2.7 AeroDR Generator Interface Unit
Status Error description Corrective actions
LED (green) does not light. AC power not supplied.
Make sure that the Ethernet cable is
connected correctly. Or, make sure the power
switch of the AeroDR Interface Unit is on.
LED (green) does not change
from ashing to a solid light.
There is a communications error.
Contact Konica Minolta technical
representatives.