5.2.11 AeroDR Battery Charger2
Status Error description Corrective actions
Status LED (green) does not
light.
AC power not supplied.
Make sure that the power cable is connected
correctly.
The power switch is turned o. Turn on the power switch.
Error LED (orange) is on. -
Contact Konica Minolta technical representa-
tives.
When the DR Detector is insert-
ed, the error LED (orange) lights.
DR Detector was inserted the wrong way.
Check the insertion orientation of the DR
Detector.
The DR Detector is not inserted correctly.
Make sure that the DR Detector is inserted
correctly.
Although the DR Detector is
inserted, charging does not start,
and the status LED (green) is out.
AC power not supplied.
Make sure that the power cable is connected
correctly.
The power switch is turned o. Turn on the power switch.
Although the DR Detector is
inserted, it cannot be registered,
and the status LED (green) is
ashing.
The Ethernet cable is disconnected.
Make sure that the Ethernet cable is connect-
ed correctly.
5.2.12 AeroDR Access Point
If a problem occurs in the AeroDR Access Point, contact Konica Minolta technical representatives.
5.2.13 Image processing controller/Images
Status Error description Corrective actions
Transverse (noise) has gotten
into all images acquired from the
DR Detector.
The wired cable is not connected properly.
Connect the wired cable spring connector unit
horizontally to the wired connector of the DR
Detector.
Operation is normal, but prob-
lems can be seen with exposure
images.
It has been occurring frequently since a cer-
tain time.
Perform calibration.
There is trouble in only 1 image.
Check the exposure method and image pro-
cessing.
Exposure cannot be conrmed in
image processing controller.
The image processing controller does not
become Ready.
Check the image processing controller start
up.
The icons of devices used on the image pro-
cessing controller are not displayed.
Conrm that the icon of the device to be used
is displayed on the system monitor screen.
When the icon of the device is not displayed,
check that the device to be used is started. Or,
check that the Ethernet cable is properly con-
nected.