83
Chapter 3
By registering the DR Detector in the image process-
ing controller using one of the following methods, the
DR Detector can be moved between exposure rooms.
Insert it into the AeroDR Battery Charger or AeroDR
Battery Charger2
Connect the wired cable to the wired connection
connector
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When using ImagePilot as the image processing
controller, it is not possible to move the DR Detector
between di󰀨erent exposure rooms.
Once it is registered in a new exposure room, it will
not be usable in the previous exposure room. When
returning to a previous exposure room, perform the
registration operation again.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3.4.1 Registration with AeroDR
Battery Charger
To register the DR Detector in the AeroDR Battery
Charger, follow the procedure below.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
AeroDR 1012HQ and AeroDR 3 1012HQ cannot be
registered with the AeroDR Battery Charger.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
AeroDR 2 1417S, AeroDR 3 1417HD and AeroDR 3
1717HD can be registered with the AeroDR Battery
Charger.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Make sure that all the devices in the destina-
tion exposure room are usable.
2
Insert the DR Detector into AeroDR Bat-
tery Charger in the destination exposure
room.
Registration process will start.
3.4 Registration and selection of the DR Detector
Exposure
side
3
Once the registration is completed, a
buzzer will sound on the AeroDR Battery
Charger, and the LED (green) will change
from slow ashing to lit.
LED (green)
4
Conrm that the DR Detector icon is dis-
played on the image processing control-
ler.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When an AeroDR Battery Charger exclusively for
charge is used, conrm that the battery charger in the
destination has a "Charge and registration" label.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
Charging will also start when the DR Detector (AeroDR
1417HQ, AeroDR 1417S, AeroDR 1717HQ, AeroDR 2
1417HQ) is inserted into the AeroDR Battery Charger,
and the AeroDR Battery Charger LED (blue) will light.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta AeroDR System and is the answer not in the manual?

Konica Minolta AeroDR System Specifications

General IconGeneral
BrandKonica Minolta
ModelAeroDR System
CategoryMedical Equipment
LanguageEnglish

Related product manuals