3.2 Operation of DR Detector
68
HINT
••••••••••••••••••••••••••••••••••••
If the DR Detector remains unused for a long time (time
can be set) it transitions to the sleep mode.
When the image processing controller is ready to ex-
pose, it recovers from the sleep mode.
••••••••••••••••••••••••••••••••••••••••••••••••••••
3.2.3 Exposure using Aero
Storage
IMPORTANT
•••••••••••••••••••••••••••••••••••••
The following DR Detector models can be used for
Aero Storage exposure: AeroDR 3 1417HD, AeroDR 3
1717HD, and AeroDR 3 1012HQ.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Starting Aero Storage Mode
Press and hold the mode LED for three seconds.
Ensure that the mode LED (blue) blinks slowly,
the information LED indicating the number of
saved images is on, and the status LED (blue) is
on.
Status LED (blue)
Information LED (white)
Mode LED
(blue)
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When DR Detector is in the following conditions, Aero
Storage mode cannot start:
It is connected to the AeroDR Battery Charger,
AeroDR Battery Charger2, and a wired cable.
DR Detector has 100 Aero Storage images saved.
DR Detector battery level has dropped to less than
10 % level.
Selection SW cannot be operated while Aero Storage
mode is active.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta AeroDR System and is the answer not in the manual?

Konica Minolta AeroDR System Specifications

General IconGeneral
BrandKonica Minolta
ModelAeroDR System
CategoryMedical Equipment
LanguageEnglish

Related product manuals