3.2 Operation of DR Detector
69
Chapter 3
Exiting Aero Storage Mode
Insert DR Detector into the AeroDR Battery Char-
ger and AeroDR Battery Charger2, or connect it
to a wired cable.
Aero Storage mode exits, and DR Detector is
connected to the image processing controller.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
You can also exit Aero Storage mode by pressing and
holding the mode LED for three seconds.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
Aero Storage mode automatically exits in the following
conditions:
DR Detector battery level drops to less than 3 %
level.
Aero Storage reaches the number of saved image
limit of 100 images.
DR Detector detects an error.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
Exposure
Exposure with Aero Storage is performed with the
following procedure.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
If you do not plan to expose immediately, exit Aero
Storage mode.
In Aero Storage, the maximum exposure time is set
to 1.8 seconds and the resolution is set to the default
(200μm). Please contact Konica Minolta technical rep-
resentatives for more information about how to change
the settings.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
1
Start Aero Storage mode.
2
Check that this device is ready to expose
images, and then prepare for the expo-
sure.
3
Push the exposure switch of the X-ray de-
vice to perform the exposure.
After exposure, images are saved to DR Detec-
tor, and the information LED displays the num-
ber of saved images.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
The DR Detector is precision equipment, and there-
fore impact or vibration during radiography or image
transfer may a󰀨ect the image quality. Be careful when
handling the DR Detector during and just after radiog-
raphy.
The number of saved images displayed in the infor-
mation LED is counted when the orange status LED
blinks slowly after the X-ray exposure.
The number of saved images displayed after an expo-
sure is the patient’s image number.
Images exposed using Aero Storage do not have
patient information. It is recommended, therefore, to
keep record so that you can associate the patient ex-
posed with the number of saved images displayed on
DR Detector.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
Exit Aero Storage mode.
Insert DR Detector into the AeroDR Battery
Charger and AeroDR Battery Charger2, or con-
nect it to a wired cable.
Aero Storage mode exits, and DR Detector is
connected to the image processing controller.
5
Import the images using the image pro-
cessing controller.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
While importing the images, do not disconnect DR De-
tector from AeroDR Battery Charger, AeroDR Battery
Charger2, or the wired cable.
When the image import process nishes, the number
of saved image in DR Detector becomes zero.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta AeroDR System and is the answer not in the manual?

Konica Minolta AeroDR System Specifications

General IconGeneral
BrandKonica Minolta
ModelAeroDR System
CategoryMedical Equipment
LanguageEnglish

Related product manuals