54
Calibration - Offsets
Adjust the sensor readings to match a known standard(s).
Current Reading (uncorrected) = raw sensor reading with no offset applied.
Offset = Value that will be added to the Current Reading (uncorrected). The
adjusted reading (uncorrected + offset) is the value reported on all screens
outside of the Sensor Calibration screen.
1. To adjust the calibration offset, touch Offset field of the desired sensor. A pop-up
will display, which allows an offset value to be adjusted for the sensor.
2. Using a known standard as reference, change the offset to match the value
reported by the standard. For example, if the Sump / Water Temp reports an
uncorrected reading of 35
o
C, and the standard reference reports 36
o
C, change
the offset to +1
o
C.
When the Offset value is changed, the current date will be saved as the Last
Cal. Date
Offset limits:
Temperature = +/- 5
o
C
Conductivity = +/- 100 µS
Fill Level = +/- 5%
Figure 6-16

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco FlaskScrubber 4121010 Series and is the answer not in the manual?

Labconco FlaskScrubber 4121010 Series Specifications

General IconGeneral
BrandLabconco
ModelFlaskScrubber 4121010 Series
CategoryDishwasher
LanguageEnglish

Related product manuals