Project SERVICE MANUAL
Date
05-2017
Version 05
39/84
5.3 Column and accelerometer calibration
After replacing always calibrate:
a) Any column.
b) Control box.
c) Any accelerometer.
Calibration of columns means:
Calibrating the speed of the columns during up or down movement.
Calibration of accelerometers means:
Calibrating the mattress platform, backrest and calfrest accelerometers in the horizontal position.
Connect the bed to the mains and activation the GO button.
5.3.1 Column and accelerometer calibration using the LCD Multiboard panel
Press the „settings“ button on the Multiboard
panel which is located on the head siderails.
1
2
3
4
5